MINMAX 10W Medical Safety Certified Power Converter

Taiwan
Introduction
Introducing the MINMAX MIW10M series - an advanced range of high-performance 10W medical-approved isolated DC-DC converters encapsulated in a DIP-24 package, meticulously crafted for medical applications. With a diverse range of 24 models supporting input voltages of 12, 24, and 48VDC, featuring a wide 2:1 input range and fixed output voltage, this series ensures adaptability to various specifications in the medical device realm.
The MIW10M series boasts an I/O isolation specified for 5000VAC with reinforced insulation, rated for a reliable 250Vrms working voltage. Advanced
features include under-voltage, overload, over-voltage, and short-circuit protection, along with no minimum load requirement, EMI emission EN 55011 class A approval, low leakage current of 2µA max, and an operating ambient temperature range from -40℃ to +90℃ without derating, achieved through high efficiency up to 88%.
Aligned with the 4th edition medical EMC standard, the MIW10M series holds medical safety approval with 2xMOPP (Means Of Patient Protection) per the 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES 60601-1, incorporating an 8mm creepage and clearance.
In adherence to ISO 14971 Medical Device Risk Management, the MIW10M series undergoes a comprehensive risk assessment process. This ensures not only compliance with high-performance standards but also alignment with the stringent safety benchmarks outlined in ISO 14971. Elevate your medical devices with the MINMAX MIW10M series - a integration of advanced technology, safety, performance, and meticulous Medical Device Risk Management Report Acquisition.
Features / strengths
►Industrial Standard DIP-24 Package
►Wide 2:1 Input Voltage Range
►Fully Regulated Output Voltage
►I/O Isolation 5000VAC with Reinforced Insulation, rated for 250Vrms
Working Voltage
►Creepage & Clearance Distance meet 8mm
►Low I/O Leakage Current < 2µA
►Operating Ambient Temp. Range -40℃ to +90℃
►No Min. Load Requirement
►Under-Voltage, Overload/Voltage and Short Circuit Protection
►EMI Emission EN 55011 Class A Approved
►Medical EMC Standard with 4th Edition of EMI EN 55011 and EMS EN
60601-1-2 Approved
►Medical Safety with 2xMOPP per 3.2 Edition of IEC/EN 60601-1 &
ANSI/AAMI ES60601-1 Approved with CE Marking
►Risk Management Report Acquisition according to ISO 14971
Specification in detail
I/O Isolation
5000VAC
Safety
ANSI/AAMI ES60601-1(2×MOPP)
Protection
Under-Voltage, Overload/Voltage and Short Circuit Protection
Supporting document

Information
Introduction
Introducing the MINMAX MIW10M series - an advanced range of high-performance 10W medical-approved isolated DC-DC converters encapsulated in a DIP-24 package, meticulously crafted for medical applications. With a diverse range of 24 models supporting input voltages of 12, 24, and 48VDC, featuring a wide 2:1 input range and fixed output voltage, this series ensures adaptability to various specifications in the medical device realm.
The MIW10M series boasts an I/O isolation specified for 5000VAC with reinforced insulation, rated for a reliable 250Vrms working voltage. Advanced
features include under-voltage, overload, over-voltage, and short-circuit protection, along with no minimum load requirement, EMI emission EN 55011 class A approval, low leakage current of 2µA max, and an operating ambient temperature range from -40℃ to +90℃ without derating, achieved through high efficiency up to 88%.
Aligned with the 4th edition medical EMC standard, the MIW10M series holds medical safety approval with 2xMOPP (Means Of Patient Protection) per the 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES 60601-1, incorporating an 8mm creepage and clearance.
In adherence to ISO 14971 Medical Device Risk Management, the MIW10M series undergoes a comprehensive risk assessment process. This ensures not only compliance with high-performance standards but also alignment with the stringent safety benchmarks outlined in ISO 14971. Elevate your medical devices with the MINMAX MIW10M series - a integration of advanced technology, safety, performance, and meticulous Medical Device Risk Management Report Acquisition.
Features / strengths
►Industrial Standard DIP-24 Package
►Wide 2:1 Input Voltage Range
►Fully Regulated Output Voltage
►I/O Isolation 5000VAC with Reinforced Insulation, rated for 250Vrms
Working Voltage
►Creepage & Clearance Distance meet 8mm
►Low I/O Leakage Current < 2µA
►Operating Ambient Temp. Range -40℃ to +90℃
►No Min. Load Requirement
►Under-Voltage, Overload/Voltage and Short Circuit Protection
►EMI Emission EN 55011 Class A Approved
►Medical EMC Standard with 4th Edition of EMI EN 55011 and EMS EN
60601-1-2 Approved
►Medical Safety with 2xMOPP per 3.2 Edition of IEC/EN 60601-1 &
ANSI/AAMI ES60601-1 Approved with CE Marking
►Risk Management Report Acquisition according to ISO 14971
Specification in detail
I/O Isolation
5000VAC
Safety
ANSI/AAMI ES60601-1(2×MOPP)
Protection
Under-Voltage, Overload/Voltage and Short Circuit Protection
Supporting document